LC-10T liquid chromatograph
The new LC-10T liquid chromatograph consists of an LC-10Tvp infusion pump and SPD-10Tvp variable-wavelength UV detector, along with an injector (manual injection valve or autosampler), chromatographic columns, and a chromatography data workstation. The chromatography data workstation is recommended to use Survey software. It utilizes a digital communication system, offers an ultra-wide dynamic range, and supports computer-based reverse control of the liquid chromatograph. It is widely applicable in a wide range of analytical fields, including pharmaceuticals, food, chemicals, and environmental analysis. The new HPLC instrument complies with GLP/GMP, FDA, and 21 CFR Part 11 regulations.
Product Details
The new LC-10T liquid chromatograph consists of an LC-10Tvp infusion pump and SPD-10Tvp variable-wavelength UV detector, along with an injector (manual injection valve or autosampler), chromatographic columns, and a chromatography data workstation. The chromatography data workstation is recommended to use Survey software. It utilizes a digital communication system, offers an ultra-wide dynamic range, and supports computer-based reverse control of the liquid chromatograph. It is widely applicable in a wide range of analytical fields, including pharmaceuticals, food, chemicals, and environmental analysis. The new HPLC system complies with GLP/GMP, FDA, and 21 CFR Part 11 regulations. ▮ Standard Configuration LC-10T
◾ LC-10Tvp High-Pressure Constant-Flow Infusion Pump (1 unit)
◾ SPD-10Tvp UV Detector (1 unit)
◾ Solvent Tray (1 unit)
◾ 8725i Manual Injection Valve (1 unit)
◾ Chromatography Workstation (1 unit)
◾ Micro-Injector (1 unit) (50μL)
◾ Injection Holder (1 unit)
(Configuration may vary slightly depending on requirements)
▮ Performance and Features LC-10T
The LC-10Tvp liquid chromatography series received a comprehensive upgrade in 2016 and features a new version of the GMP-compliant chromatography workstation, including survey functionality, electronic signatures, user rights, audit tracking, and system compatibility. Exterior and Craftsmanship Upgrades:
◾ Door Design Changes
◾ More Optimized Piping Design
◾ Improved Cooling System
◾ Replaceable Tray Design
◾ Improved Conductive Silicone Button Design
◾ Changed Instrument Signal Communication Method
▮ Technological Enhancement of the LC-10T
LC-10Tvp High-Pressure Constant-Flow Infusion Pump:
◾ Infusion Method: Micro-Volume Series Dual-Plunger
◾ Maximum Infusion Pressure: 6000 psi
◾ Flow Setting Range: 0.001–9.999 ml/min (Adjustable in 0.001 ml/min Steps)
◾ Flow Setting Error: ≤0.5%
◾ Flow Stability Error: ≤0.2% RSD
◾ Pressure Pulsation: ±1% (0-100%, Water/Acetone Water Gradient)
◾ Pump Tightness: Less than 0.1 MPa Flow rate 1 mL/min, pressure drop less than 5 MPa
◾ Time program function: Yes
◾ Dimensions: W300 x H135 x D435 mm
SPD-10Tvp UV-Vis variable wavelength detector:
◾ Wavelength range: 190 nm to 700 nm
◾ Wavelength indication error: ≤ ±1 nm
◾ Wavelength repeatability error: ≤ ±0.1 nm
◾ Dynamic noise: ≤ ±0.75 × 10-5 AU (methanol, 1 mL/min, 254 nm, 20°C)
◾ Static Noise: ≤±0.5×10-5 AU (empty cell, 1 second response time, 20°C)
◾ Dynamic Baseline Drift: ≤±1×10-4 AU/h (methanol, 1 mL/min, 254 nm, 20°C)
◾ Static Baseline Drift: ≤0.5×10-4 AU/h (empty cell, 1 second response time, 20°C)
◾ Linear Range: ≥104
◾ Minimum Detectable Concentration: ≤4×10-9 g/mL (naphthalene/methanol solution)
◾ Qualitative Repeatability: RSD ≤0.1%
◾ Quantitative Repeatability: RSD ≤0.5%
◾ Spectral Bandwidth: 6 nm
◾ Flow Cell Volume: 10 μL
◾ Optical Path Length: 10 mm
◾ Time Programming: Yes
◾ Dimensions: W300*H135*D435 mm